Sign in →

Test Code RPDESO Rapidly Progressive Dementia Evaluation, Spinal Fluid

Important Note

Not available for outreach clients to order. Internal orders only.

Regional Centers: Do not open bag or perform any testing on these specimens. Send specimens to Marshfield for handling/processing.

Marshfield Lab Staff: Do not open bag or perform any testing on these specimens. Follow the procedure "Handling of CSF Samples from Patients with Suspected or Confirmed Cretuzfeld-Jakob Disease" found in Document Control System.

Additional Codes

Mayo Test Code: RPDE


Ordering Guidance


In individuals with a high clinical suspicion of Alzheimer disease, order ADEVL / Alzheimer Disease Evaluation, Spinal Fluid.

 

This test can only be performed on specimens collected and transported in polypropylene tubes. If this test is ordered and a polystyrene tube is received, it will be canceled and automatically reordered by the laboratory as CJDE / Creutzfeldt-Jakob Disease Evaluation, Spinal Fluid.

 

For cases where there is high suspicion of human prion disease supported by clinical or paraclinical magnetic resonance imaging features, order CJDE / Creutzfeldt-Jakob Disease Evaluation, Spinal Fluid.

Early in the disease course, or in atypical cases, the disease progression may be slower and include significant clinical overlap (dementia, rigidity, myoclonus) with other potential causes of rapidly progressive dementia, including Alzheimer disease. In the latter case, it would be more appropriate to order this test.



Specimen Required


Supplies: CJD/RPD Evaluation Kit (T966)

Container/Tube:

Preferred: 2 Sarstedt CSF False Bottom Tubes 63.614.625 (2.5 mL)

Acceptable: Sarstedt 72.703.600 (1.5 mL) or Sarstedt 72.694.600 (2 mL)

Specimen Volume: 2 tubes; each containing 1.5 to 2.5 mL

Collection Instructions:

1. Perform lumbar puncture and discard the first 1 to 2 mL of cerebrospinal fluid (CSF).

2. Collect two tubes of CSF directly into an acceptable collection tube until the tube is at least 50% full.

3. Send CSF specimen in original collection tube. Do not aliquot.

Note: Polystyrene collection tubes are not acceptable. Exposure of CSF to polystyrene tubes may result in falsely low Abeta42 concentrations.

The Alzheimer's Association consensus protocol for handling of CSF for clinical measurements of Abeta42 and tau recommends using the drip method for CSF collection and directly collecting into a low-bind polypropylene tube. Although some clinicians prefer the syringe pull method due to speed of collection, the drip method reduces the risk of Abeta42 binding to the plastic of any syringe used.

4. Collection instructions can also be found on Spinal Fluid Specimen Collection Instructions for Creutzfeldt-Jakob Disease and Rapidly Progressive Dementia Evaluations (T974).


Useful For

Evaluation of individuals presenting with rapidly progressive dementia of uncertain disease etiology and a differential diagnosis of Creutzfeldt-Jakob disease and rapidly progressive Alzheimer disease

Profile Information

Test ID Reporting Name Available Separately Always Performed
RPDEI RPD Eval Interp, CSF No Yes
RTQPC RT-QuIC Prion, CSF No Yes
TTPTQ t-Tau/p-Tau No Yes
ADRTQ Alzheimer's Disease Evaluation, CSF No Yes

Method Name

RPDEI: Medical Interpretation

RTQPC: Real-Time Quaking-Induced Conversion (RT-QuIC)

TTPTQ: Calculation

ADRTQ: Electrochemiluminescent Immunoassay (ECLIA)

Reporting Name

Rapid Progress Dementia Eval, CSF

Specimen Type

CSF

Specimen Minimum Volume

See Specimen Required

Specimen Stability Information

Specimen Type Temperature Time Special Container
CSF Frozen (preferred) 28 days BlueTop SARSTEDT
  Refrigerated  14 days BlueTop SARSTEDT
  Ambient  12 hours BlueTop SARSTEDT

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject
Discolored CSF Reject

Reference Values

RT-QuIC PRION, CSF:

Negative

 

t-TAU/p-TAU:

≤18

 

p-TAU/ABETA 42:

≤0.028

 

BETA-AMYLOID (1-42) (Abeta42):

>834 pg/mL

 

TOTAL TAU:

≤238 pg/mL (Alzheimer disease)

≤393 pg/mL (Creutzfeldt-Jakob disease)

 

PHOSPHORYLATED TAU 181:

≤21.6 pg/mL

Cautions

These test results should be interpreted in the appropriate clinical context along with other clinical and paraclinical findings. Only through neuropathological assessment of brain tissue can a definitive diagnosis of sporadic prion disease be established.

 

Some molecular subtypes of prion protein have been reported to have lower detectability by the real-time quaking-induced conversion (RT-QuIC) assay.

 

Even small quantities of blood in CSF can result in false-negative RT-QuIC results.

 

The presence of fluorescent substances may interfere with testing and prevent the accurate interpretation of the RT-QuIC assay.

 

Careful consideration of the differential diagnosis is advised when RT-QuIC test results are unexpectedly negative. Repeat testing with RT-QuIC may be warranted if there is high suspicion of prion disease. A small subset of initially negative cases by RT-QuIC may become positive as the disease progresses. However, a small proportion of patients with definitive prion disease may be persistently negative by RT-QuIC. False-negative RT-QuIC results are most often encountered in cases of genetic prion disease, such as fatal familial insomnia and Gerstmann-Straussler-Scheinker, and in atypical sporadic prion disease subtypes that have slower indolent disease progression.

 

Improper specimen handling or interindividual differences in overall concentration of Abeta peptide production may yield an abnormally low Abeta42 in the context of a normal p-Tau181/Abeta42 ratio. Results should be interpreted in combination with other clinical information.

 

Exposure of cerebrospinal fluid to polystyrene tubes can reduce concentrations of the amyloid Abeta42 by as much as 20% to 50% due to adherence of the sticky amyloid protein to polystyrene tube surface material, potentially altering clinical interpretation, including the p-Tau181/Abeta 42 ratio. P-Tau181 and total Tau protein do not substantially adhere to polystyrene collection tubes.

 

Failure to adhere to the specimen collection instructions provided may result in falsely low Abeta42 concentrations and potential misdiagnosis of Alzheimer disease.

 

In rare cases, some individuals can develop antibodies to mouse or other animal antibodies (often referred to as human anti-mouse antibodies [HAMA] or heterophile antibodies), which may cause interference in some immunoassays. The presence of antibodies to streptavidin or ruthenium can also rarely occur and may interfere in this assay. Caution should be used in interpretation of results, and the laboratory should be alerted if the result does not correlate with the clinical presentation.

Day(s) Performed

Monday through Friday, Sunday

Report Available

3 to 8 days

Specimen Retention Time

12 months

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

0584U

82234

84393

84394

LOINC Code Information

Test ID Test Order Name Order LOINC Value
RPDE Rapid Progress Dementia Eval, CSF 104134-2

 

Result ID Test Result Name Result LOINC Value
PTABQ p-Tau/Abeta42 41027-4
620307 RT-QuIC Prion, CSF 101662-5
TTPTQ t-Tau/p-Tau 101752-4
620377 RPD Eval Interp, CSF 69048-7
ADINQ AD Interpretation 69048-7
AB42Q Abeta42 33203-1
TTAUQ Total-Tau 30160-6
PTAUQ Phospho-Tau(181P) 72260-3